TABLE 3 - GUIDELINES AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC INTERFERENCE RESISTANCE
Interference
IEC 60601
Resistance
EN 60601-1-2
(Immunity) Test
Conducted RF
3 V eff
EN 61000-4-6
From 150 kHz to 80 MHz
Radiated RF
3 V/m
EN 61000-4-3
From 80 MHz to 2.5 GHz
AQUOLAB PROFESSIONAL
Disconnect the device from the supply mains before any maintenance activity
ORDINARY CLEANING
Dry the water tank (f.) with a cloth after every use to prevent limestone scale build-up.
ORDINARY MAINTENANCE – ANTI-SCALING TREATMENT
Aquolab Professional alerts you to the need for ordinary maintenance using the symbol (i) (see
the "Home Page" figure). By clicking it, you will open the "Maintenance" page.
Remember that, at least once a month or when Aquolab gives you this signal, you need to run
a de-scaling cycle with the specific de-scaling solution supplied in vials (Aquolab De-scaling
solution) which you can purchase directly from the www.aquolab.com website. The de-scaling
cycle allows you to remove scale build-up resulting from calcium and magnesium salts in
water. This will maintain continuing device efficiency. To avoid damaging the device, do not use
products other than the Aquolab de-scaling kit.
Follow these steps to correctly use the Aquolab De-scaling solution:
Shake the vial before using it.
Pour the contents into the dental water jet tank and fill it with water.
Switch the device on and let it dispense water for approximately 15 sec.
Wait 5 minutes and repeat this operation for a further 15 sec.
Repeat the operation until the tank is completely empty.
Rinse the tank by filling it with water only and letting it flow out at regular intervals until
the tank is completely empty.
To reset the "Ordinary maintenance" alarm and make Aquolab Professional ready for use
again, simply press the reset button on the "Maintenance" page.
(IMMUNITY)
Compliance
Level
3 V eff
Portable and mobile RF communications devices should be used no closer to "AQUOLAB
From 150 kHz
PROFESSIONAL" or any of its components (including cables) than the recommended separation
to 80 MHz
distance determined using the equation applicable to the frequency of the transmitter.
Where P is the maximum output power rating of the transmitter in watts (W) according to the
information provided by the transmitter manufacturer and d is the recommended separation
3 V/m
From 80 MHz to
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, must
2.5 GHz
Interference may occur in the vicinity of equipment marked with the following symbol:
®
Electromagnetic Environment – Guidelines
Recommended separation distance:
d = 1.2√P 150 kHz to 80 MHz
Recommended separation distance:
d = 1.2√P 80 MHz to 800 MHz
Recommended separation distance:
d = 2.3 √P 800 MHz to 2.5 GHz
distance in metres (m).
be less than the compliance level in each frequency range.
MAINTENANCE