4.7.1 Test program types

Selecting the "test" sub-menu accesses the list of test cycles that autoclave can run:
- Vacuum Test
- 134°C Helix Test
- 121°C Helix Test
- 134°C Bowie&Dick Test

4.7.2 Vacuum test (leakage test)

To run this function, the chamber must be cold and have no instruments in it.
4.7.3 Bowie & dick test
To run this test, the chamber must have no instruments in it. The Bowie & Dick "package" used for the porous
load test must be positioned on the lower tray.

4.7.4 Helix test

To run this test, the chamber must have no instruments in it. The instrument used to test narrow hollow loads must
be positioned on the lower tray.

4.7.5 Test diagrams

4.7.5

Diagrammi test

Nisea Diagramma134°C Helix e Bowie & Dick Test
Nisea Diagramma 121°C Helix Test
Pressure
Pressure
Time
Time
Nisea Diagramma Vuoto Test
39 39
Pressure
Time
38
39 39
60