8. Battery Replacement

Follow the procedure below and replace batteries with
new ones (type IEC LR03 1.5V).
• Unscrew the battery door, e.g. with a coin.
• Pull out the Battery door and replace the batteries. Insert
• Re-assemble battery door.
• Confirm that the Battery door case is properly locked prior

9.Technical data

Voltage range:
LED Nominal voltage:
ELV indication
Response time:
Peak current:
Measurement Duty:
Internal battery
consumption:
Single-pole phase test
voltage range:
Phase rotation test:
Continuity test:
Battery:
Temperature:
Humidity:
Altitude:
Overvoltage:
Standard:
Pollution:
Protection:
1760PT.indd 9
Remove the probes from any testing point, when opening
the Battery case. Batteries are dead when the continuity
test with both test probes connected cannot be done
anymore.
new batteries according to the engraving on the Battery
door.
to measurements.
Battery Replacement
12...690V AC (16...400Hz), DC(±)
12/24/50/120/230/400/690V, AC
(16...400Hz), DC(±)
LED tolerances according to
EN61243-3
LED, >50 V AC/DC
< 1s at 100% of each
nominal voltage
Is<3.5mA (at 690V)
30s ON (operation time), 240s OFF
(recovery time)
Approx. 80mA
100...690V AC (50/60Hz)
120...400V earth-to-phase,
AC 50/60Hz
Detection range 0...500kΩ + 50%
3V (IEC LR03 1.5V x 2)
-15...55°C operation; -20...70°C
storage, No condensation
Max 85% RH
up to 2000m
CAT III/1000V, CAT IV 600V
IEC 61243-3:2014
degree 2
IP 64
9