Inspect all connections to make certain there are no
leaks.
•
To remove the water line from the Cavitron Plus System,
turn off the dental office water supply. Disconnect the
water supply line from the dental office water supply. If
a quick-disconnect connector is attached to the end of
the hose, relieve the water pressure by pressing the tip
of the connector in an appropriate container and allow
water to drain. To remove the hose from the system,
push on the outer ring of the system's water inlet and
gently pull out the water line.
6.7 Foot Control Battery
Installation/replacement
•
Turn foot control over and using a Philips screwdriver
carefully remove battery cover screw and battery cover.
If applicable, remove used batteries and install two new
"AA" batteries as shown. Do not depress foot control
while installing batteries.
•
The communication light will blink for approximately
two seconds to indicate the foot control's ability to
communicate with the unit. If the light does not blink,
check the batteries. If the batteries are good and
the light doesn't blink, a communications error may
exist. Re-establish communication with Foot Control
Synchronization procedure Section 6.10.
•
The remote frequency communication can be bypassed
using the auxillary foot control cable. Refer to Section
10.2 Technical Support and Repair for further action.
•
Replace the battery cover and screw and hand tighten
with Philips screwdriver.
Press ring to release
water supply tube.
Look for blinking
communications
light.
•
Remove batteries if foot control is to be stored for an
extended period of time.
6.8 Foot Control Synchronization
The wireless foot control supplied with your system has been
factory synchronized with the base unit. Should a replacement
foot control be necessary, synchronization will be required
prior to system operation. Perform the following steps to
synchronize the foot control with the base unit.
1.
Turn the Main Power switch located at the center
front underside of the system to the OFF (O) position.
2.
Install a new set of "AA" batteries into the foot control
(See Section 6.7) Leave the battery cover of the
foot control open so the red push button is
accessible.
3.
Maintain a distance of no more than 10 feet between
the base unit and foot control during the
synchronization process.
4.
Turn the Main Power switch to the ON (I) position
and wait for the Diagnostic Display graphics to light
(refer to Section 7.2).
5 . While all graphics are lit, press the Purge button,
located on the Diagnostic Display. The graphics will
begin to blink in a sequential pattern, representing
the synchronization mode. This mode will last 5 to 6
seconds.
6.
During this mode, press the red button located in the
battery compartment of the foot control. This will
complete the synchronization process.
7.
Synchronization is successful when all graphics blink
at the same time.
8.
To verify proper communication, press the foot
control to the Boost position (foot control fully
pressed – 2
position) and ensure the Boost graphic
nd
on base unit lights.
9.
Replace battery cover and the screw.
7