t200 - Sensor expired
t400 - Pressure sensor trouble
t800 - PCBA not tested
If the error mode repeats frequently, check for continuous power and proper voltage. If power is not the problem and a unit has
repeating error conditions, it may need to be returned to Macurco for service, per these User Instructions.
If the error mode indicates "Sensor expired" see section
4 Maintenance
The CD-6"x" is low maintenance. The unit uses a long life NDIR sensor that has a 15-year life expectancy (in normal conditions). The
CD-6H uses an automated background calibration program to set the clean air level on a regular basis. The CD-6H will maintain accuracy
if it is exposed to the "clean air reference value" (this reference value is the lowest concentration to which the sensor is exposed) at
least once per week.
Note: This applies when used in the typical indoor ambient air with weekly unoccupied periods.
All maintenance and repair of products manufactured by Macurco are to be performed at the appropriate Macurco manufacturing facility.
Macurco does not sanction any third-party repair facilities.
Note: There is not a field calibration procedure for the Macurco CD-6H.
4.1 End-of-life Signal
The CD-6"x" has a long life, the non-replaceable infrared sensor (IR Sensor). Fifteen (15) years after the CD-6"x" is installed the sensor
end-of-life signal will be activated indicating that the CD-6"x" has reached the end of its typical usable life. The end-of-life signal will cause
an error code t200 "Sensor expired". See section
"ENTER/TEST" button or by temporarily dropping power to the unit. The end-of-life signal provides the user an opportunity to test and/or
calibrate the sensor (CD-6MC only) assuring that it is still performing within acceptable parameters though the sensor is nearing the end
of its expected life. The silence function will continue to be available for 29 days after the CD-6"x" initiates the initial end-of-life signal.
After this 29-day period, the CD-6"x" can no longer be silenced and the unit must be replaced.
Do not disassemble unit or attempt to repair or modify any component of this instrument. This instrument contains no user serviceable
parts and substitution of components may impair intrinsic safety, which may adversely affect product performance and result in
sickness or death.
Avoid the use of harsh cleaning materials, abrasives and other organic solvents. Such materials may permanently scratch the surfaces
and damage the display window, labels, sensor or instrument housing. High voltage terminals (100-240VAC) are located within this
detector, presenting a hazard to service technicians. Only qualified technicians should open the detector case and service the internal
circuits. Ensure power is removed from the detector prior to cleaning the unit. Failure to do so may result in serious injury or death.
4.2 Cleaning
Cleaning of the external surfaces is best carried out using a damp cloth with a mild detergent or soap. Use a vacuum cleaner with a soft
brush to remove dust or contamination under the cover. Do not blow out the sensor with compressed air.
4.1 End-of-life Signal
3.6.2 Error
Codes. The end-of-life signal can be silenced for 48 hours by pressing the
WARNING
CAUTION
16
of these User Instructions.