X-Mind Novus e
CRA-2
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transients/bursts
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions and
voltage variations
on power supply
lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Reviewed: Lavonen Maria 2012-12-05 13:19
Approved: Savinen Antti Juhani 2013-02-01 15:29
IEC 60601 test level
Compliance level
6 kV contact
6 kV contact
8 kV air
8 kV air
2 kV for power
2 kV for power supply
supply lines
lines
1 kV for input/output
1 kV for input/output
lines
lines
1 kV differential
1 kV differential mode
mode
2 kV common mode
2 kV common mode
<5 % U
<5 % U
T
(>95 % dip in U
)
(>95 % dip in U
T
for 0.5 cycle
for 0.5 cycle
40 % U
40 % U
T
(60 % dip in U
)
(60 % dip in U
T
for 5 cycles
for 5 cycles
70 % U
70 % U
T
(30 % dip in U
)
(30 % dip in U
T
for 25 cycles
for 25 cycles
<5 % U
<5 % U
T
(>95 % dip in U
)
(>95 % dip in U
T
for 5 sec
for 5 sec
3 A/m
3 A/m
Apéndice A. Información técnica
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
T
)
should be that of a
T
typical commercial or
hospital environment. If
user of the device
T
)
requires continued
T
operation during power
mains interruptions, it is
recommended that the
T
)
device be powered from
T
an uninterruptible power
supply or a battery.
T
)
T
Power frequency
magnetic field should be
at levels characteristic
of a typical location in a
typical commercial or
hospital environment.
A-13
D508773, 2