Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration – electromagnetic immunity
The BM 47 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 47 should assure that
it is used in such an environment.
IMMUNITY test
Electrostatic discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The BM 47 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 47 should assure that
it is used in such an environment.
IMMUNITY test
IEC 60601
test level
IEC 60601 test level
± 6 kV contact
± 8 kV air
3 A/m
Compliance
Electromagnetic environment – guidance
level
Compliance level
± 6 kV contact
± 8 kV air
3 A/m
100
Electromagnetic environment –
guidance
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.