Damping of the analogue output
The time delay can be set in 5 steps via a second menu. When the menu is selected, the current damp-
ing level is displayed.
The blue ring flashes and the 5 LEDs indi-
cate the respective setting.
Menu item
Tier 1 (min.)
Tier 2
Tier 3
Tier 4
Tier 5 (max.)
26
The probe must be disconnected from the supply
voltage.
Adjustment with EasyTeach-Button (ETF):
Push and hold the EasyTeach-Button when switch-
ing on the supply voltage and the menu for the damp-
ing is activated. The Menu goes through the various
damping tiers in an infinite loop. The menu starts
with the previous adjusted tier and changes every
5 seconds. The desired damping is set by releasing
the EasyTeach-Button after the specified time
Adjustment with EasyTeach by Wire (ETW):
The menu for the damping starts by activating the
EasyTeach-Connection (Teach wire / Teach pin with
the supply voltage UB+) when switching on the sup-
ply voltage. By continuing to hold the EasyTeach-
Connection actively the menu goes through the
various tiers of the damping in an infinite loop. The
menu starts with the previous adjusted tier and
changes every 5 seconds. The desired damping is
set by disconnecting the wire after the specified time.
Optical representation
Blue ring flashes
+ LED Ref. is on
Blue ring flashes
+ LED 25% is on
Blue ring flashes
+ LED 50% is on
Blue ring flashes
+ LED 75% is on
Blue ring flashes
+ LED 100% is on