Blood Pressure Measurement Error Messages
E06
Inflation time too long.
Causes: No cuff is connected.
Solution:
E09
Incorrect cuff
Causes:
Solution:
E10
Arrhythmia
Causes:
Solution:
E11
Artefact
Cause:
Solution:
E12
Excessive pressure in the cuff or measurement time too long
Cause:
Solution:
E14
Leak in the cuff or in the connector cable
Cause:
Solution:
Comment: Errors E10 and E11 cannot always be clearly classified; for some arrhythmias
an E11 may be displayed.

Instructions for Care

Device:
The device and the sensor should be cleaned with a soft, clean cloth.. Please use ethyl
alcohol or isopropyl alcohol for deep cleaning. Other agents, in particular substances
that contain acetone, will damage the housing or the display.
Cuffs
Cleaning:
After removing the pad, the nylon Velcro covers can be wiped with a moist cloth or, just
like all other cuffs, be washed with soap and cold water. Should you opt for the latter, rinse
the cuff with clean water and air dry it. Wipe the pad and the tube with a moist cloth.
Disinfection:
After removing the pad, the cuff covers can be washed with a disinfectant in cold water
and then air dried.
The pad and tube can be wiped with a bit of ethanol and a cotton cloth.
ATTENTION!
Nylon Velcro cuffs may not be ironed! Never subject the cuffs to intense solar ra-
diation!
30
The cuff is too loose on the arm.
The cuff is not tight.
Place the cuff again and start a new measurement.
Adult cuff in neonate mode.
Neonate cuff in adult mode.
Use the correct cuff.
The device determined implausible readings during the last mea-
surement.
Take another measurement.
The device has detected too many motion artefacts.
Take another measurement.
Pressure was applied on the cuff during the measurement.
Do not touch the cuff and tube during the measurement.
The device recorded an unexpected decrease in pressure during
the measurement.
Use a different cuff and check the cable to the device.