Body Temperature; Technical Specifications - Riester ri-vital Instrucciones De Uso

Ocultar thumbs Ver también para ri-vital:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Do not touch the cuff cover nor the pad with a pointy object as this could cause
damage!
Body Temperature Measurement
The device can be optionally equipped with a holder for the ri-thermo® N thermometer.
This holder enables easy storage of the thermometer and provides the required protective
cover for the measuring head.
For more information about using the device and important instructions regarding measu-
ring precision, please refer to the operating manual for the ri-thermo® N.

Technical Specifications

General
Mains operation:
Mains adapter:
Capacity:
Protection class:
Moisture proofing:
Patient module:
Battery:
Operating time:
Charging:
Display:
Temperatures:
Humidity:
Mechanical stability:
Dimensions:
Weight:
Housing:
Note for clinicians: Information regarding the range of the top wavelengths and maximum
optical capacity of the light emitted by the pulsoximeter can be useful to clinicians who are
applying photodynamic therapy.
Pulsoximeter
Sensors:
Measurement range:
Wavelength for the
measurement:
220/230 V, 50 Hz, through mains adapter
12V,
4 W max.
II
IPX1
BF for SpO2
Li-ion battery, 7.4 Volt 450 mAh
about 500 measurements
About 8 hours
High-contrast LCD with illumination
Storage:
-5 to 60° C
Operation:
10 to 40 ° C
0...95 % relative humidity, non-condensing
Vibration test pursuant to EN ISO 9919:2005
Drop test pursuant to EN ISO 9919:2005
246 mm x 167 mm x 232 mm with handle + ri-thermo
203 mm x 167 mm x 232 mm with handle
(L x W x H)
1.650g with ri-thermo
1.500g without ri-thermo
Break-proof, easy-to-clean PC + ABS plastic
All PEARL sensors
SpO2
0 - 100 %
Pulse
30 - 249 1/min
660 nm/880 nm
®
N
®
N
®
N