Technical Measurement Check - Riester ri-vital Instrucciones De Uso

Ocultar thumbs Ver también para ri-vital:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Velcro cuff children
Velcro cuff adults, thin arms
Velcro cuff adults
Velcro cuff adults, large arms
Mobile stand with basket
ri-thermo® N
Package with 100 disposable probe covers
Package with 1000 disposable probe covers

Technical Measurement Check

Germany:
Pursuant to the Ordinance on Operating Medical Devices (Medizinprodukte-Betreiberver-
ordung, MPBetreibV), a technical check must be conducted in two-year intervals. The check
can only be conducted by the manufacturer, the appropriate authorities or persons who
comply with the requirements of MPBetreibV §6.
European Union outside of Germany:
The respective legal provisions apply for all EU countries outside of Germany.
1. Procedure for Checking Compliance with the Error Threshold of the
Reading Display of the Cuff Pressure
1.1 Test equipment
a)
Solid metal container with a capacity of 500 ml +/- 5%.
b)
Calibrated control manometer with a maximum permissible error of measurement
of 1.0 mmHg (0.1 kPa).
c)
Pressure generator, e.g. pump ball with discharge valve.
d)
T pieces and tubes.
1.2 Procedure
a)
Replace the device's cuff with the metal container (see 1.1a). Connect the cali-
brated control manometer (see 1.1b) with the cuff connector (see Fig. 1) by
means of a T piece and tube (see 1.1d) with the cuff connector (see Fig. 1).
Connect the additional pump ball (see 1.1c) to the pneumatic system with the
help of another T piece.
In order to get to the calibration mode, turn the device off and press the 12, 13
and 15 buttons simultaneously until "100" appears on the display; when you let
go of the buttons, "E01" will be displayed and the device will be in calibration
mode for 170s.
b)
Conduct the test in pressure ranges of at most 50 mmHg between 0 mmHg and
the highest value of the display range.
Exceeding of the overpressure can take place at any desired level. The test can
be continued after decreasing the overpressure to the desired value.
(article number LF129)
(article number LF120)
(article number LF107)
(article number LF122)
(article number 12660)
(article number 12704)
(article number 12705)
33

Publicidad

Tabla de contenido
loading

Tabla de contenido