the batteries are almost empty
In such cases, repeat the measurement. Ensure you do not
move or speak.
If necessary, re-insert or replace the batteries.
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic)
lie outside the limits specified in the section "Technical
specifications", the technical alarm will appear on the dis-
play indicating either "Hi" or "Lo". In such cases, you should
seek medical assistance and check the accuracy of your
procedure.
The limit values for the technical alarm are factory set and
cannot be adjusted or deactivated. These alarm limit val-
ues are accorded second priority under the standard IEC
60601-1-8.
The technical alarm is a non-locking alarm and must not be
reset. The signal shown on the display will disappear auto-
matically after about 8 seconds.

7. Cleaning and care

Clean your blood pressure monitor carefully using a
slightly damp cloth only.
Do not use any detergents or solvents.
Never hold the instrument under water as otherwise liquid
can penetrate and damage the instrument.
Never place any heavy objects on the instrument.
.
19

8. Technical details

Model no.
BC 44
Measurement
Oscillometric, non-invasive blood pres-
method
sure measurement on the wrist
Measurement
Cuff pressure 0 – 300 mmHg,
range
systolic 60 – 260 mmHg,
diastolic 40 –199 mmHg,
Pulse 40 –180 beats/minute
Display accuracy Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement
Max. permissible standard deviation
inaccuracy
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Dimensions
L 87,8 mm x W 65,8 mm x H 28,6 mm
Weight
Approx. 98 g (without batteries)
Cuff size
140 to 195 mm
Permissible oper-
+ 10 °C to + 40 °C, ≤ 90 % relative air
ating conditions
humidity (non-condensing)
Permissible stor-
- 20 °C to + 55 °C, ≤ 90 % relative air hu-
age conditions
midity, 800 –1050 hPa ambient pressure
Power supply
2 x 1,5 V
AAA batteries