Operation

NOTICE
The Dräger X-am 5000 must be equipped with an
NiMH power supply unit T4 (order no. 83 18 704). The
use of other power supply units is not permitted!
When switching on the Dräger X-zone 5000, Dräger
recommends wearing hearing protection or using the
alarm-attenuation ring (order no. 83 20 110) as the
acoustic alarm is activated for a brief period.

Switching on the instrument

STANDBY mode
z Release the lock on
the holder.
z Remove the holder.
z Hold down the
key on the
OK
Dräger X-zone 5000
for approx. 3 seconds.
– The battery LED will be
lit in green, red/green
or red (refer to "Status
LED overview" on
page 46) depending on
the battery capacity.
– The Dräger X-
zone 5000 is in the
STANDBY mode.
z Place the Dräger X-am 5000 into the device
receptacle.
NOTICE
The clip on the Dräger X-am 5000 must be closed.
z Place the holder onto the housing.
z Fasten the lock at the holder.
CAUTION
ON mode
When switching on the Dräger X-zone 5000, Dräger
recommends wearing hearing protection or using the
alarm-attenuation ring (order no. 83 20 110) as the
acoustic alarm is activated for a brief period.
z Switch on the Dräger X-am 5000 in accordance with
the Instructions for Use.
– The visual and the audible alarm will be activated for
a short time.
– The Dräger X-am 5000 switches to the X-zone mode
(refer to "X-zone mode:" on page 29).
– The visual and acoustic life signal (green LED ring
and single tone) is issued depending on the
configuration (1 - 60 seconds; default setting:
2 seconds).
– The Dräger X-zone 5000 is in the ON mode and
ready for operation.
– In ON mode, the alarm signals of the Dräger X-am
5000 are evaluated, processed, and, if applicable,
transmitted to other Dräger X-zone 5000 devices.
X-zone mode:
– The battery symbol
is replaced by the X-zone mode symbol
Dräger X-zone 5000.
– The Dräger X-am 5000 is supplied via the Dräger
X-zone 5000 battery.
– If the activated Dräger X-am 5000 is removed from
the Dräger X-zone 5000, the device will exit X-zone
mode after max. 10 seconds.
CAUTION
on the Dräger X-am 5000
on the
29