Further Information - Heyl Testomat 2000 PO4 Manual De Instrucciones

Fotómetro de procesos para fosfato 0 –10 mg/l
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Error
Current interface functions incorrectly
Unit is not functioning, even though it is
switched on
No display
48

Further information

Possible causes
- Incorrect measured value at the output or
no power supplied
- Fuse F9, F5 or F2 (240 V: F1) defective
- Power switch defective
- Ribbon cable at display circuit board or
base circuit board is loose
- Error at display circuit board or base
circuit board
Tripping of a protective circuit
After a protective circuit (fuse) has been tripped, attempt to eliminate
the cause of malfunctioning (e.g. replace a defective valve) before
reactivating the protective circuit. Frequent triggering is always due to
power overload which, in certain circumstances, may also damage to
the instrument.
Malfunctioning/Repairing a defective instrument
The repair of a defective instrument – irrespective of the warranty
period – is only possible when the instrument is dismantled and
returned to us with a description of the error. Furthermore, please
inform us of the measured medium. Before you return the instrument
for repair work, remove the bottles and ensure that the measuring
chamber has been flushed out and is empty.
Error messages/Troubleshooting
Remedies
 Replace fuse F7
 Replace the interface circuit board
 Replace fuses
 Replace power switch
 Reconnect ribbon cable
 Replace display or base circuit board