Scope of delivery
®
1x Testomat 2000
PO4
2 screw caps with hole. Used for reagent bottles
1x operating instructions
Performance specifications
The process photometer Testomat 2000
chemical online monitor for monitoring the phosphate content from 0
to 10 mg/l (ppm) using the photometric analysis principle. Analysis is
carried out by adding two reagents and the analysis result is
displayed after a reaction time of ten minutes.
Measuring range: 0 – 10 mg/l with a resolution of:
- 0.1 mg/l in the range of 0 – 7 mg/l
- 0.25 mg/l in the range of 7 – 10 mg/l
Limit of detection: 0.1 or 0.25 mg/l depending on range
Accuracy: < 10% at end of measuring range
Freely selectable unit in ppm or mg/l
Analysis initiation:
- Automatic interval operation
(Interval pause adjustable from 0 to 255 minutes)
- External control
- Volume controlled (water meter)
Two independent limit values with hysteresis (1, 2 or 3 bad
analyses) and adjustable switch functions
Monitoring of two measuring points (change-over via external
solenoid valves)
Internal error documentation
Programmable service address
Programmable maintenance interval for a maintenance
request
Container volume can be varied between bottle with 500 ml
capacity and container with 20/5 litre capacity (a set with two
screw caps with suction lance is available to connect both
containers)
optional
Interface (0/4-20 mA or 0/2-10 V) or
Interface RS 232 for a log printer or
SD card data logger
Scope of delivery
®
PO4 is a robust wet-
9