CAUTION
(C)
(D)
(A)
(B)
18
Commissioning
Handling of reagents/indicators
Adhere to the respective safety data sheet!
Trouble-free operation of Testomat 2000
when using Heyl Testomat 2000
Connect the reagents according to the description for your container
size.
Inserting large containers (5/20 litre canister)
Open the lower housing door by pulling on the right side.
Put the canisters under the device.
Remove the caps from the canisters.
Insert one suction lance into each canister.
Note the length of the suction lances!
The shorter suction lance (Part No. 40536) is for the 5 litre
canister (B) (Reagent B), the longer suction lance (Item No. 40535)
is for the 20 litre canister (A) (Reagent A).
Hand-tighten the screw caps of the suction lances onto the
canisters.
Guide the suction lances through the cable ducts (C) on the right
side of the housing wall.
Insert the suction hose by hand into the hose connector of the
insert (D).
Pay attention to ensure the correct assignment of reagents A (20 l)
and B (5 l) to the dosing pumps: A = left, B = right
In the basic program under Program values -> Volume, select
20/5l canister.
Inserting the reagent bottles (500 ml)
Open the lower housing door by pulling on the right-hand side.
Remove the cap from the indicator bottle.
Remove the plastic bag from inside the lower housing door. The
plastic bag contains the screw cap with hole
for the screw cap for each bottle.
Place the insert
into the bottle .
Screw the screw cap with hole
Place the bottle underneath the dosing pump. Observe correct
allocation of the reagents A and B to the dosing pumps: A = left,
B = right.
Push the intake hose
insert
.
®
PO4 is only guaranteed
®
PO4 reagents!
and the insert
hand-tight onto the bottle
hand-tight onto the hose connector of the
Commissioning
.