3) Keep pressing the [Mode] key until F4 oFF is displayed. 
4) Press the [‐>0/T<‐] key to view the current settings. 
5) Press [Mode] to change the settings (0, 5, 10, 20 and 30 minutes). 
6) Press [‐>0/T<‐] to store a particular setting. The display will return to F4 oFF 
7) When F4 oFF is displayed press the [Print] key to return to weighing or press 
[Mode] to go to the next function. 
 
6.5
SELECTING
M
APPROVED 
You can select if the internal mass or an external mass is to be used to calibrate the 
balance.  
1) To  set  this  parameter  press  the  [Mode]  key  during  self‐checking  test  when 
turning on the balance. 
2) After a few seconds, the display will show the first function F1 UNT.  
3) Keep pressing the [Mode] key until F5 IEC is displayed. 
4) Press the [‐>0/T<‐] key to view the current settings.  
5) To  change  the  settings  press  the  [Mode]  key  to  change  from  INt  (internal 
calibration) or E (external calibration) 
6) Press [‐>0/T<‐] to store a particular setting. The display will return to F5 IEC 
7) When F5 IEC is displayed press the [Print] key to return to weighing or press 
[Mode] to go to the next function. 
 
 
 
 
© Adam Equipment Company 2018
THE
INTERNAL
OR
 
 
 
ODELS
EXTERNAL
CALIBRATION
 
 
(N
 
OT POSSIBLE ON 
19