3
Maintenance and service
Replacing the lamp
Caution Halogen lamps are pressurized to provide
maximum efficiency and illumination. Mistreatment may
cause shattering. Protect lamp surface against abrasions
and scratches. Use protective eyeglasses when
handling. Dispose of lamp with care.
Caution Only use Welch Allyn lamp REF 06500-U.
Caution Lamps may be hot. Allow the lamps to cool
before removing.
Caution Avoid removing the lamp for reasons other
than replacement.
It is necessary to re-calibrate the white balance on the Digital
MacroView after replacing the lamp.
15