1
. Terms for safe operation
st
User's better use the device in accordance to the design of it; it may result in
equipment damage or personal injury if used incorrectly or inappropriately:
● Not used in accordance with design requirements;
● Operators and maintainers use it without training;
● Users change the contents of the design without authorization;
● Users don't pay attention or understand the safe-using rules.
Low-speed centrifuge should read and understand the
operation way and rules of safe-use in this manual.
You must totally enforce the following rules to avoid accidents:
The "Manual" is part of the accessories of the "Low-speed table-top centrifuge", it
must be put together with the machine for the operator to check. The Low-speed
centrifuge is designed for clinical medicine, biology, chemistry, genetic engineering,
immunology, etc. The density of the separated sample cannot exceed 1.2g/cm ³ under
the maximum speed, when the density of the sample exceeds 1.2g/cm ³, the maximum
speed of the rotor must be reduced accordingly.
Please make sure that nobody or any hazardous materials are within a 30cm
distance from the centrifuge, and no materials block the vent of the centrifuge when the
Low-speed centrifuge is running (separating the sample or the rotor is rotating).
The centrifuge will injure the operator or other staff or damage the centrifuge and
the samples inside if the user doesn't follow the following safety measures when
operating the Low-speed Table-top centrifuge:
● The centrifuge can't be used in a corrosive environment or if there is possibility of
explosion because the design of the centrifuge is neither anti-corrosion, nor is
explosion proof;
● Do not use following materials in the centrifuge:
- Flammable and explosive materials;
- Strong chemical agents;
Versión 08 Enero 2018
Each person involved in the use or maintenance of the
Pág. - 29 -