Aquatec® ORCA / ORCA F / ORCA XL
Harmonic
emissions
Class A
IEC 61000-3-2
Voltage fluctuations
/flicker emissions
Complies
IEC 61000-3-3

10.3 Electromagnetic Immunity

Guidance and manufacturer´s declaration
This product is intended for use in the electromagnetic environment specified below. The customer or the user of this product
should assure that it is used in such an environment.
Immunity test
Test / Compliance level
Electrostatic
± 8kV contact
discharge (ESD)
± 2 kV, ± 4 kV, ± 8 kV,
IEC 61000-4-2
15 kV air
± 2 kV for power supply
lines; 100 kHz repetition
Electrostatic
frequency
transient / burst
± 1 kV for input / output
IEC 61000-4-4
lines; 100 kHz repetition
frequency
Surge
± 1 kV line to line
IEC 61000-4-5
± 2 kV line to earth
28
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment.