Basics
4.2
Initial setup of station
1. Switch on the station, see chapter 4.1 on page 17.
2. Log in with the preconfigured "admin" user
(user name: admin, password: 123456), see chapter 4.5
on page 18.
3. Configure the test gas inlet, see chapter 4.6 on page 18.
4. Change language if necessary:
a. Select
> System configuration > Language.
b. Selected the required language.
c. Confirm selection with OK.
5. Set date and time where necessary:
a. Select
> System configuration > Date & time.
b. Make the desired settings.
c. Confirm settings with OK.
4.3
Touchscreen display
The buttons on the touchscreen display change dynamically
depending on the task being executed. To execute an action,
select the corresponding icon on the display.
Press the
key on the master at any time to access the start
screen.
4.4
Start and test screens
The keys on the start and test screen change dynamically
depending on the log-in status, single-mode status and the
number of modules in use. For more information see the X-
dock 5300/6300/6600 Technical Manual.
4.4.1
Symbols
Menu
Confirm
Cancel
Back
Log user in or out
18
Select this button to access the
menu.
Select this button to confirm an
input or function.
Select this button to cancel an
input or function.
Select this button to access the
previous screen.
Select this button to log users in
or out. The number in the
symbol indicates the permission
level (see chapter 6 on
page 23).
4.5
Log user in or out
NOTICE
i
i
A User-ID is required to log in. This must be created
beforehand by the Administrator.
A user with administrator rights is created by default:
User name: admin
Password: 123456
NOTICE
i
i
Dräger recommends changing the admin password
after initial start-up.
To log in a user:
1. Select
.
a. Select
.
b. Select the desired user name from the list.
or
a. Select Select user.
b. Enter desired user name.
2. Enter the password and confirm with
NOTICE
i
i
When entering the user name, 3 already saved user
names will automatically be suggested and displayed.
For quick selection, please select the desired user
name.
To log out the current user:
1. Select
.
Information about the current user will be displayed.
2. Select
.
The current user will be logged out
4.6
Configuring the test gas inlet
WARNING
!
The gas concentrations entered must be identical to
the specifications on the gas cylinder used. Incorrect
details will result in faulty measurement results.
To configure a gas inlet:
1. Select
> Station gas configuration.
An overview of the test gas connections is displayed.
.
Dräger X-dock 5300/6300/6600