Switching The Station On Or Off; Initial Setup Of Station; Touchscreen Display; Start And Test Screens - Dräger X-dock 5300 Instrucciones De Uso

Ocultar thumbs Ver también para X-dock 5300:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 51
Dräger recommends selecting the "Purge" test option to purge
the module with fresh air after performance of a test.
Dräger recommends not storing an X-am 8000 with PID in the
module with the cover closed for a longer period of time, to
avoid drifting of the PID.
For precise measurements with the PID LC, Dräger
recommends repeating the zero adjustment on the sampling
site with the X-am 8000 using an activated carbon pre-tube.
The span calibration does not need to be repeated.
If an X-am 8000 with a PID LC has been stored for a longer
period of time in the X-dock, the calibration must be checked
prior to using the gas detector.
4
Basics
4.1

Switching the station on or off

NOTICE
i
i
If no action has occurred for 10 minutes, any user
logged in will be logged off automatically. After
45 minutes, the screensaver will be activated.
To switch the station on:
Press and hold the
1 second.
The following information is displayed during the switch-on
process:
Software version number
To switch the station off:
Press and hold the
3 seconds.
The station switches off.
Standby mode:
Standby mode is activated after approx. 10 minutes with no
activity on the station (input via touchscreen or opening/
closing of a module cover).
When the station switches over to standby mode, any user
logged in will be logged off automatically. The user will
need to log back in again when switching back to operating
mode.
The touchscreen is switched off in standby mode.
The charging function of the X-am 125+ and X-am 8000+
modules with charging function is not affected by standby
mode. The charging process is not interrupted.
To switch to operating mode:
Tap the function key or
touch the touchscreen or
open or close a module flap.
4.2

Initial setup of station

NOTICE
i
i
During the initial setup, the maintenance station starts
in X-dock Manager mode (XDM mode). To change the
mode see Chapter 6.4 on Page 29.
Dräger X-dock 5300/6300/6600
key on the master for approx.
key on the master for approx.
1. Switch on the station, see Chapter 4.1 on Page 23.
2. Log in with the preconfigured "admin" user
(user name: admin, password: 123456), see Chapter 4.5
on Page 24.
3. Configure the test gas inlet, see Chapter 4.6 on Page 24.
4. Change language if necessary:
a. Select
> System configuration > Language.
b. Select desired language.
c. Confirm selection with OK.
5. Set date and time where necessary:
> System configuration > Date & time.
a. Select
b. Make the desired settings.
c. Confirm settings with OK.
4.3

Touchscreen display

The buttons on the touchscreen display change dynamically
depending on the task being executed. To execute an action,
select the corresponding icon on the display.
Press the
key on the master at any time to access the start
screen.
4.4

Start and test screens

The keys on the start and test screen change dynamically
depending on the log-in status, single-mode status and the
number of modules in use. For more information see the
X-dock 5300/6300/6600 Technical Manual.
4.4.1
Symbols
Menu
Select this button to access the
menu.
Confirm
Select this button to confirm an input
or function.
Cancel
Select this button to cancel an input
or function.
Back
Select this button to access the
previous screen.
Log user in or
Select this button to log users in or
out
out. The number in the symbol
indicates the permission level (see
Chapter 6 on Page 28).
XDM server
Connected
connection
XDM server
Not connected
connection
Basics
23

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

X-dock 6300X-dock 6600

Tabla de contenido