Switching The Station On Or Off; Initial Setup Of Station; Touchscreen Display; Start And Test Screens - Dräger X-dock 5300 Instrucciones De Uso

Ocultar thumbs Ver también para X-dock 5300:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 43
WARNING
!
Risk of personal injury!
Impurities of the ambient air may result in faulty
measurements.
If you use the internal pump to supply fresh air through
the fresh air inlet you need to ensure that the ambient
air is devoid of interfering substances.
7. Connect the power pack.
Station with up to 3 modules: Power pack 24 V / 1.33 A
Station with 4 to 10 modules: Power pack 24 V / 6.25 A
The entire system is supplied with power via the master.
NOTICE
i
i
Dräger recommends the use of Dräger test gas
cylinders and Dräger pressure regulator valves (see
chapter 9 on page 28). Alternatively there is the option
of using a suitable pressure regulator valve with
0.5 bar outlet pressure and >3 L/min flow rate.
Dräger recommends connecting an exhaust hose
(max. 10 m long) to the exhaust outlet to discharge the
test gas into the open air.
4
Basics
4.1

Switching the station on or off

NOTICE
i
i
If no action has occurred for 10 minutes, the station
switches automatically to standby mode.
To switch the station on:
Press and hold the
1 second.
The following information is displayed during the switch-on
process:
Software version number
To switch the station off:
Press and hold the
3 seconds.
The station switches off.
Standby mode:
Standby mode is activated after approx. 10 minutes with no
activity on the station (input via touchscreen or opening/
closing of a module cover).
When the station switches over to standby mode, any user
logged in will be logged off automatically. The user will
need to log back in again when switching back to operating
mode.
The touchscreen is switched off in standby mode.
The charging function of X-am 125+ modules with charging
function is not affected by standby mode. The charging
process is not interrupted.
Dräger X-dock 5300/6300/6600
key on the master for approx.
key on the master for approx.
To switch to operating mode:
Tap the function key or
touch the touchscreen or
open or close a module flap.
4.2

Initial setup of station

1. Switch on the station, see chapter 4.1 on page 19.
2. Log in with the preconfigured "admin" user
(user name: admin, password: 123456), see chapter 4.5
on page 20.
3. Configure the test gas inlet, see chapter 4.6 on page 20.
4. Change language if necessary:
a. Select
> System configuration > Language.
b. Select desired language.
c. Confirm selection with OK.
5. Set date and time where necessary:
a. Select
> System configuration > Date & time.
b. Make the desired settings.
c. Confirm settings with OK.
4.3

Touchscreen display

The buttons on the touchscreen display change dynamically
depending on the task being executed. To execute an action,
select the corresponding icon on the display.
Press the
key on the master at any time to access the start
screen.
4.4

Start and test screens

The keys on the start and test screen change dynamically
depending on the log-in status, single-mode status and the
number of modules in use. For more information see the X-
dock 5300/6300/6600 Technical Manual.
4.4.1
Symbols
Menu
Confirm
Cancel
Back
Log user in or out
Basics
Select this button to access the
menu.
Select this button to confirm an
input or function.
Select this button to cancel an
input or function.
Select this button to access the
previous screen.
Select this button to log users in
or out. The number in the
symbol indicates the permission
level (see chapter 6 on
page 24).
19

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

X-dock 6300X-dock 6600

Tabla de contenido