6 . 9 R es ume t r ea tmen t a f ter o pen lea d
Press "oK" after checking electrodes and cables connections. Treatment screen is displayed. Amplitudes
have to be readjusted manually. A cable breakage can be checked by pressing the cable's pins against
one another while increasing the amplitude for the corresponding channel to 11 mA. If the amplitude
now drops to 0.0 mA and starts flashing, the cable needs to be replaced.
note: Never increase the amplitude above 20 mA when you check for cable breaks, since this can
damage the stimulator.
6 . 1 0 P r o gr am e nd
When the program is finished, the last used programs are selected for the next time the
stimulator is turned on.
The stimulator turns off after 5 minutes of inactivity to spare battery.