Problem
Display does not come on
Weak stimulation with loaded
batteries
stimulation stops with loaded
batteries
stimulation weakens within
minutes of starting treatment
with loaded batteries
stimulation is uncomfortable
Possible Cause
Low battery
Electrodes dried out, lost their
adhesive power and have no
adequate connection to the
skin.
Acupuncture mode is wrongly
selected
Electrode placement
Poor electrode contact
Damaged or worn electrodes
Cabel or contacts damaged
This is a normal body adaptive
process
Amplitude (intensity) is too
high
Electrodes are too close
together
Damaged or worn electrodes
Ensure proper program is being
used
Solution
Charge battery
Replace electrode
Unselect acupuncture mode
Make the electrodes at least 5
cm (2 inches) apart
Reapply electrodes, secure
firmly. Electrodes must be a
minimum of 5 cm apart.
Replace
Replace
Increase the amplitude
(intensity) if required.
Decrease amplitude (intensity).
Reposition the electrodes.
Electrodes must be a minimum
of 5 cm (2 inches) apart.
Replace
A. Refer to the programs
description
B. Contact clinician if discomfort
persists