Beurer BC 16 Manual De Instrucciones página 21

Tensiómetro
Ocultar thumbs Ver también para BC 16:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 34
9. Specifications
Model no.
BC 16
Measurement
Oscillometric, non-invasive blood
method
pressure measurement on the wrist
Measurement range
Cuff pressure 0 – 299 mmHg,
systolic 30 – 260 mmHg,
diastolic 30 – 260 mmHg,
Pulse 40 – 199 beats/minute
Display accuracy
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement
Max. permissible standard deviation
inaccuracy
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory
2 x 50 memory spaces
Dimensions
L 82 mm x W 69 mm x H 28 mm
Weight
Approx. 120 g (without batteries)
Cuff size
135 to 220 mm
Permissible operating
+10 °C to +40 °C, 40 – 85 % relative
conditions
air humidity (non-condensing)
Permissible storage
-10 °C to +60 °C, 10 – 90 % relative
conditions
air humidity, 800-1050 hPa ambient
pressure
Power supply
2 x 1,5 V
Battery life
For approx. 300 measurements,
depending on the blood pressure
level and/or pump pressure
Accessories
Instruction for use, 2 x 1.5 V AA bat-
teries, storage box
Classification
Internal supply, IPX0, no AP or
APG, continuous operation, type BF
applied part
Technical information is subject to change without notifica-
tion to allow for updates.
This unit is in line with European Standard EN 60601-1-2
and is subject to particular precautions with regard to
electromagnetic compatibility (EMC). Please note that
portable and mobile HF communication systems may
interfere with this unit. More details can be requested from
the stated Customer Service address or found at the end
of the instructions for use.
This device is in line with the EU Medical Devices Direc-
tive 93/42/EC, the "Medizinproduktegesetz" (German
Medical Devices Act) and the standards EN 1060-1
(non-invasive sphygmomanometers, Part 1: General
requirements), EN 1060-3 (non-invasive sphygmoma-
nometers, Part 3: Supplementary requirements for elec-
AA batteries