4.3.2 Influence on frequency
Values to be added to the reference conditions.
NOTE: linear interpolation between each specified value (see also the graphs in chapter 7 APPENDIX).
Frequency
Current to be
in A AC
measured
in % of In
Error to be added
4.3.3

Environmental conditions

I I I I I Indoor use
I I I I I Altitude: ≤ 2 000 m
I I I I I Transportation altitude: ≤ 12 000 m
I I I I I Temperature influence: < 0.1% / 10°C
I I I I I Climatic conditions:
% HR
100
90
80
70
60
50
40
30
20
10
0
-50
-40
-30
-20
4.4

Conformity to international standards

4.4.1
Electrical safety (in accordance with NF EN 61010-2-032)
- Dual insulation
- Pollution degree 2
- Installation category III
- Working voltage: 600 V rms
4.4.2
Compatibilité électromagnétique
I Emission: meets EN 50081-1
I Immunity: meets EN 50082-2
An influence ≤ 10% of the output signal is accepted for the test of immunity to radiated fields in the
110 - 120 MHz and 170 - 180 MHz frequency bands
2 A
25 A
0.4%
5%
4.5% +1 mV
4.5%
-10
0
10
20
30
40
14
65 Hz to 1000 Hz
100 A
250 A
500 A
20%
50%
3%
2%
°C
50
60
70
80
90
600 A
In
120%
1%
2%