STOMACHER
7.3.2

EU and UK Products

Model
Cat No.
Dimensions
Weight
Supply Voltage
Supply Frequency
Plug Top Fuse
Mains Inlet Fuses*
Transformer Fuse*
PCB Fuse*
Insulation
Installation
Pollution
Duty Cycle
Rated Load
Disposable Bag Size
Minimum Recommended
Capacity
Maximum Recommended
Capacity
Maximum Recommended
Sample Temperature
Recommended Operating
Ambient Temperature Range
Recommended Operating
Relative Humidity Range
Timer Settings
No Load Paddle Speeds
(@ nominal 20°C)
Paddle Clearance
* Only these or equivalent UL-approved fuses should be used.
Issue: 3.2
®
3500 AND 3500 BIOWASHER
USER INSTRUCTIONS
®
Sttomacher
3500 and 3500 Biowasher
3500: 3500/000/EU and 3500/000/UK
3500 Biowasher: 3500/001/EU and 3500/001/UK
480mm wide x 480mm deep x 445mm high (unpacked)
36kg (unpacked)
220-250V
50-60Hz
F5A (UK cord set only)
T1.0A 250V 5x20mm, breaking capacity: 35A @ 250V
T250mA 250V 5x20mm, breaking capacity: 35A @ 250V
F2.5A 250V 5x20mm, breaking capacity: 100A @ 250V
Class 1 (Earthed)
Over Voltage Category II
Pollution Degree 2
Continuous
120W
380mm x 508mm
400ml
3500ml
60°C
10-35°C
10-89%
0 seconds – 99 minutes 59 seconds (±5%)
75rpm to 180rpm in 5rpm increments (±5%)
0 - 20mm