8.2

Control Key Functions

The control panel, located on the top of the Stomacher
1.
POWER switch: powers up the Stomacher
2.
TEMPERATURE CONTROLLER: allows the desired temperature setting to be
selected between 20 and 50°C. The upper (green) display indicates the actual
temperature and the lower (orange) display indicates the set or target temperature.
The set temperature can be adjusted by: - Press the Η (STAR) and ▲ (ARROW UP)
keys to increase the set temperature. Press the Η and ▼ (ARROW DOWN) keys to
decrease the set temperature. Once set this data is stored until reset and retained
when the Stomacher is switched off.
3.
HEATER ON/OFF switch: the temperature controller will automatically adjust the heater
when the HEATER switch is in the on position. The LED adjacent to the switch will
illuminate to indicate when the heater is switched on.
4.
CYCLE TIME CONTROLLER: allows the desired cycle run time setting to be selected
between 0 and 99 minutes, 59 seconds. The upper display indicates the cycle run time
remaining and counts down. The lower display indicates the set cycle run time. The cycle
run time can be adjusted by pressing the appropriate green button on the controller to
increment each digit in turn. Once set this data is stored until reset even if the
Stomacher
5.
The CYCLE switch enables the operator to select either of the following: -
TIMED
the selected time.
STOP
switch does not control the heater.
CONTINUOUS
6.
SPEED selector switch: enables LOW, MEDIUM or HIGH speed to be selected by
turning the switch to the appropriate position.
Issue: 3.2
STOMACHER
USER INSTRUCTIONS
®
is switched off.
-
The Stomacher
-
The Stomacher
-
The Stomacher
®
3500 Thermo
®
, has the following functions: -
®
and is illuminated when on.
®
runs at the selected speed for the duration of
®
will not run, but is still powered up. This
®
runs at the selected speed continuously.