Weigh Averaging

Purpose
Use this program to determine weights under unstable ambient conditions.
In this program, the balance/scale calculates the weight as the average value
from a defined number of individual weighing operations. These weighing
operations are also known as "subweighing operations" or "subweighs."
Available Features
– The measured result displayed is the arithmetic mean shown in the selected
weight unit; a triangle indicates that this is a calculated value
– You can set the number of subweighing operations performed in the operating
menu: See "Configuring the Balance/Scale"
– Press [F] for at least 2 sec. to display the preset number of subweighing
operations
– Press [F] to toggle the display between the calculated result and the weight
Factory Settings
Number of subweighs for averaging: 10 (3. 3. 2)
Preparation
§ Configure the Weigh Averaging application in the operating menu:
See "Configuring the Balance/Scale"
Menu code 2. 1.12 Averaging
– Number of subweighs for weigh averaging:
5 subweighs
3. 3. 1
10 subweighs
3. 3. 2
20 subweighs
3. 3. 3
50 subweighs
3. 3. 4
100 subweighs
3. 3. 5
See also "Configuring the Balance/Scale"