4.
Turn the mode selector switch to | (On).
5.
Rotate the suction control knob fully counter-clockwise (decrease).
6.
Clamp the connective tubing. The gauge needle should not move.
With the tubing still clamped, increase the suction to 35 cmH
7.
8.
Slowly open and close the clamped tubing to create various flow rates through
the regulator. Check that the suction level is maintained when the tubing is
clamped.
9.
Clamp the connective tubing.
10. Rotate the suction control knob fully clockwise (increase) to check that the
relief valve opens (begins to flow) at 55 cmH
kPa). As suction is increased, the relief valve should not allow levels of more
than 60 cmH
11. Reduce the suction to zero and set the mode selector switch to O (Off).
Patient Setup
1.
Make sure the Pre-Use Checkout Procedure has been performed.
7
Figure 7
Setup with reusable collection system
2.
Clamp the connective tubing and turn the mode
selector switch to | (On)
Clamp
6700-0358-000 (Rev. 9) 08/2020
O (5.9 kPa).
2
O ± 5 cmH
O (5.4 kPa ± 0.5
2
2
O (3.4 kPa).
2
7