CD-11 Indicator

2.6.4 Setup Menu

The CD-11 Indicator SEtuP Menu must be entered the first time the Indicator is used to set the scale base
parameters to match the Indicator. Do not attempt to calibrate the Indicator before initially setting up the
SEtuP Menu. All other menus should be entered and set up the first time the Indicator is used.
Procedure
With the Indicator ON, press and hold the G/N/T/ Menu button until MEnU is
displayed. When the G/N/T/ Menu button is released, CAL is displayed if the
CAL jumper on the PC board is in place. When the CAL jumper is removed, the
Indicator will not permit calibration. This jumper should be in place initially.
Press the PRINT/ Units button, SEtuP is displayed.
Press the G/N/T/ Menu button, LFtOFF is displayed. Legal for trade selections
are:
'ON' - LFT (Legal for Trade) is ON
'OFF' - LFT is OFF.
'CAn' - LFT is set for Canada
Press the PRINT/ Units button until desired LFT setting is reached.
Press G/N/T/ Menu button, 0 2 is displayed. This is the Zero setting. Selections
are:
2%:
zero operating range is -2% to +2%.
18%: zero operating range is -2% to +18%,
100%: zero operating range is -2% to +100%.
NOTE: : : : : If LFT is ON, only 2% and 18% are available.
Press the PRINT/ Units button until desired zero setting is reached.
Press the G/N/T/ Menu button, CAL Un kg is displayed. This is the calibration
unit setting. Selections are:
'lb' - calibration unit is lb
'kg' - calibration unit is kg.
Press the PRINT/ Units button until desired calibration unit setting is reached.