SJX SERIES BALANCES
4.3 Setup Menu
Enter this menu to set balance parameters.
Reset:
Auxiliary Graduation:
Filter:
Auto Zero Tracking:
Stable:
Backlight:
Auto Tare:
Auto Off:
End Setup:
Note: Bold always represents factory default value.
Reset [r r eset]
Reset the Setup menu to factory defaults.
NO
= not reset
YES
= reset
Auxiliary Graduation [Au.Grad]
Set whether the weight is displayed with the auxiliary graduation when the Security Switch is in
the locked position.
ON
= the auxiliary graduation is displayed within brackets.
OFF
= the auxiliary graduation is not displayed.
Note:
This menu item is only included in models SJX323N/E and SJX1502N/E.
The setting affects kilograms, grams, carats and pennyweights in model SJX323N/E and
kilograms, grams and pennyweights in model SJX1502N/E.
If the scale will be used in direct sales to the public, the OFF setting must be selected.
Filter [FiltEr]
Set the amount of signal filtering.
LOW
MED
HI
Auto Zero Tracking [AZt]
Set the automatic zero tracking functionality.
OFF
0.5d
1d
3d
Stable Range [Stable]
Set the amount the reading can vary while the stability symbol remains on.
0.5d
1d
2d
5d
Find Quality Products Online at:
no, yes
On, Off
Low, Med, High
off, 0.5d, 1d, 3d
0.5d, 1d, 2d, 5d
off, on, auto
off, on, on-acc
off, 1, 5, 10
Exit menu
= less stability, faster stabilization time
= normal stability, stabilization time
= greater stability, slower stabilization time
= disabled
= the display will maintain zero until a change of 0.5 divisions per second has
been exceeded.
= the display will maintain zero until a change of 1 divisions per second has been
exceeded.
= the display will maintain zero until a change of 3 divisions per second has been
exceeded.
= 0.5 balance division
= 1 balance division
= 2 balance division
= 5 balance division