4.6. Filters

Removing the nose cone and adapter exposes a hydrophobic filter
(0.45 micron) located at the top of the nose piece. The filter protects the
XL from accidental over-aspirations and we strongly recommend to use
only the original Biohit filters. Both sterile and non-sterile filters are
available.
Cat. No
712912 Filter for Biohit Proline XL, non-sterile
712913 Filter for Biohit Proline XL, sterile
Note: The XL can not be operated without a filter. Wet filters, due to
overfilling will slow or stop fluid movement. If overfilling occurs, change the
filter immediately by simply pulling it out and placing a new filter, with the
blue side upwards, into the nose piece.

5. PROGRAMMING THE XL

Programming is done using the six-button keyboard and the LC-display.
There are four different operating modes, offering special functions and
varying speeds for your selection.

5.1. Mode Selection and Mode Recall

1. Press
2. Press
operation on the selected mode.
Note: A mode can only be activated when the piston is in its home
position (arrow up sign is lit).

5.2. Setting Speeds

1. Press
2. Press
displayed (5 = Fast and 1 = Slow).
3. Press
current dispensing speed.
4. Press
displayed (5 = Fast and 1 = Slow).
5. Press
Product
repeatedly to view the modes available.
when the desired mode is displayed. The XL is ready for
to display the current aspiration speed.
or
until the desired aspiration speed is
to confirm the speed selection. The display shows the
or
until desired dispensing speed is
to confirm the speed selection.
Qty.
5 pcs
1 pcs
9