Maintenance; Troubleshooting; Decommissioning; Repair - ProMinent DULCOTEST ICT 1 Instrucciones De Servicio

Sensor de conductividad inductivo
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

ProMinent
®
Commissioning / Maintenance / Troubleshooting /
Select "install. factor setting?" in the setting menu and restore
the installation factor noted earlier.
Next, restore the temperature coefficient α of the water sample
used
It is highly recommended that you check the conductivity of the
calibration solution with suitable reference metering equipment
to be sure that the conductivity of the calibration solution has not
changed due to carry-over of liquid or evaporation.
9

Maintenance

Since it is electrode-free and is designed with good flow charac-
teristics the sensor is very low maintenance.
Remove deposits from the sensor regularly to maintain a reliable
measurement operation.
NOTE
Deposits can be prevented from forming in most cases. Install
the sensor in flowing media and make sure that the sensor tip
is in line with the direction of the flow.
10

Troubleshooting

See Operating instruction manual of the DULCOMETER
controller.
11

Decommissioning

Detach the electrical connection of the sensor from the
®
DULCOMETER
D1C.
12

Repair

The sensor must be returned to the factory for repair. For this
purpose, clean it and return to us in its original packaging with a
completed complaint form.
Decommissioning / Repair
®
D1C
37

Publicidad

Tabla de contenido
loading

Tabla de contenido