parameter
volume
NSA pause*
(minutes)
CPR Coaching
compressions/
ventilations†
CPR Coaching
ventilations
shock series
protocol pause*
(minutes)
*If the Trainer is configured for CPR Coaching with ventilations, and the user selects CPR Coaching by pressing the i-button, the default pause
and CPR Coaching settings (2 minutes pause, 30 compressions/2 ventilations) result in 5 cycles of CPR Coaching.The number of cycles varies
for other settings. For all settings, CPR Coaching begins and ends with compressions.
† The CPR Coaching compressions/ventilations selection (switch 4) is applicable only if the CPR Coaching ventilations (switch 5) is set to OFF
(with ventilations). Otherwise, the CPR Coaching compressions will be continuous, without breaks for ventilations.
Page 17
J
switch #
1
OFF:
ON:
2 & 3
switch #2
OFF
ON
OFF
ON
4
OFF:
ON:
5
OFF:
ON:
6
OFF:
ON:
7 & 8
switch #7
OFF
ON
OFF
ON
setting (default = OFF)
high
low
switch #3
setting
OFF
2.0
OFF
1.0
ON
1.5
ON
3.0
adult
30 compressions/2 ventilations
pediatric
30 compressions/2 ventilations
adult
30 compressions/2 ventilations
pediatric
15 compressions/2 ventilations
with ventilations
without ventilations
1
3
switch #8
setting
OFF
2.0
OFF
1.0
ON
1.5
ON
3.0
17