pH 330i/340i
Resetting to default settings
Measurement
parameters
Configuration
parameters
Resetting to default settings
You can reset the measurement parameters and the configuration to the de-
livery status separately from one another (initialization).
The following measurement parameters can be reset to the delivery status:
pH/ORP (pH Inl)
Measuring mode
Slope
Asymmetry
Manual temperature input
Resolution (pH display)
pH
-59.16 mV/pH
OmV
25 °C
0.01
The following configuration parameters (Inl) can be reset to the delivery sta-
tus:
Baud rate
Display during pH calibration
Interval 1
(automatically saved)
Interval 2
(for data transmission)
4800
Buffer nominal value
OFF
OFF
Procedure of the reset:
• To switch it off, press <CAL> while pressing the <RUN/ENTER> key.
• To reset the measurement parameters with <A> <T>, select YES and
confirm with <RUN/ENTER>, or
• Continue to the configuration parameters without resetting {Inl} with the
<RUN/ENTER> key.
• After the configuration parameters, Inl changes the instrument to the last
active measurement mode.
13