pH 330i/340i
Measurement
parameters
Configuration
parameters
pH
i
n i
n
o
All manuals and user guides at all-guides.com
Resetting to default settings
You can reset the measurement parameters and the configuration to the de-
livery status separately from one another (initialization).
The following measurement parameters can be reset to the delivery status:
pH/ORP (pH InI)
Measuring mode
Slope
Asymmetry
Manual temperature input
Resolution (pH display)
The following configuration parameters (InI) can be reset to the delivery sta-
tus:
Baud rate
Display during pH calibration
Interval 1
(automatically saved)
Interval 2
(for data transmission)
Procedure of the reset:
l To switch it off, press <CAL> while pressing the <RUN/ENTER> key.
l To reset the measurement parameters with < > < >, select YES and
confirm with <RUN/ENTER>, or
l Continue to the configuration parameters without resetting (InI) with the
<RUN/ENTER> key.
l After the configuration parameters, InI changes the instrument to the last
active measurement mode.
Resetting to default settings
pH
-59.16 mV/pH
0 mV
25 °C
0.01
4800
Buffer nominal value
OFF
OFF
13