3.1 - Cleaning: Irrigator compressor body
For component (5) only use a damp cloth with antibacterial detergent (non-abrasive and solvents
free).
The nasal douche has an aspiration valve (5c) which is to be cleaned when dirty. To access the valve
lift the lid of the battery compartment and remove the valve by lifting it with a ballpoint pen (fig
H); then wash the valve with water and detergent. Before reassembling, make sure that the valve is
completely dry.
ATTENTION:
The compressor body of the douche is not protected against the penetration of liquids; therefore do
not wash it with running water or immersion.The compressor body of the douche is not protected
against the penetration of liquids; therefore do not wash it with running water or immersion.
3.2 - Sanification: Clean the components (1A-1B-1C-2,3,4) choosing one of the following methods.
Method A: Wash the components with hot drinking water (about 40°C) with a delicate dish deter-
gent (non-abrasive).
Method B: through immersion in a solution composed of 50% water and 50 % white vinegar. Then
rinse well with hot drinking water (about 40°C);
Drying: After washing or sanitizing the components, shake them well and place them on a paper
napkin or dry them with hot air (for example with a blow dryer). Reassemble the nasal douche as
as
shown in the picture (see fig A).
3.3 - Disinfection: The accessories that can be disinfected are (1A-1B-1C-2,3,4)
Method A: The disinfection procedure described in this paragraph is to be performed before using
the accessories and is effective only if all the prescriptions herein are met and all the components
4),
involved are previously sanitized.
The disinfectant to be used must be of electrolytic oxidizing chlorine type, (active ingredient: so-
dium hypochlorite), specific for disinfection and is available in all pharmacies.
h
Procedure:
- Fill a container, suitable to contain all the individual components to be disinfected, with a solution of
on
drinking water and disinfectant, while respecting the proportions indicated on the packaging of the
e
disinfectant itself.
- Completely immerse each individual component in the solution, taking care to avoid the formation
of air bubbles in contact with the components. Leave the components immersed for the period of
time indicated on the packaging of the disinfectant, and associated with the concentration chosen
4c)
to prepare the solution.
sal
- Remove the components now disinfected and rinse thoroughly with lukewarm drinking water.
- Dispose of the solution as recommended by the disinfectant manufacturer.
Method B: boil in water for 10 minutes; it is advisable to use demineralized water or distilled water
in order to avoid chalky deposits.
After disinfection shake the accessories vigorously and place them on a paper towel, or alternatively, dry
he
them with a jet of hot air (e.g. using a hair dryer).
3.4 - Sterilization:
Sterilization intended for clinical or hospital use is the responsibility of the same structure that it per-
forms and must be performed according to the instructions in this manual.
The accessories that can be sterilized are (2-3-4)
The sterilization procedure described in this paragraph is effective only if all the prescriptions
gs
herein are met, all the components involved are previously sanitized and the procedure itself is
validated according to ISO 17665-1 rule.
Unit: Steam sterilizer with fractionated vacuum and overpressure according to EN 13060 rule.
Procedure: Package each component to be treated in sterile barrier system or packaging complying
with EN 11607 rule. Place the packed components in the steam sterilizer. Run the sterilization cycle
according to the instructions manual of the device, selecting a temperature of 134° C and a duration
of 10 minutes.
Storage: Store sterilized components as per instructions of the sterile barrier package or system chosen.
ADDITIONAL CLEANING INFORMATION of the atomization chamber (with the exception of the
compressor body).
In order to avoid deposits that may clog the atomizer and the atomization chamber and consequently
affect the operation of the nasal irrigator, it is recommended that cleaning and sanitizing is performed
after each use following the instructions for use contained in this instruction manual included with the
nasal irrigator. After rinsing accurately the parts under warm tap water (40°C approx.), shake the wet
9