6 Supervisor menu

The supervisor menu consists of the following blocks:
F1 – scale settings
F5 – terminal settings
F6 – exit menu

6.1 Entering Supervisor menu

➜ In gross mode, press and hold
➜ Enter password
SEtUP appears in the display.
➜ Press
.
F1 appears in the display. All parameters can be modified.
Note on certified weighing systems (OIML or NTEP)
On certified weighing systems, the parameters F1, F5.1 and F5.4 are locked.
In order to change these parameters, proceed as follows:
1. Switch off the weighing terminal and open it.
2. Press button S1 on the mainboard and switch on weighing terminal.
SEtUp is displayed, and all parameters can be modified.
3. When configuration is finished, seal the weighing terminal.
6.2 Operating the Supervisor menu
Operating the Supervisor menu means the same as operating the Operator menu, see page
15.
6.3 Block F1 – Scale
Factory settings are printed in bold letters.
F1.1 – Approval
no approval
no
approval according to OIML
OIML
approval according to NTEP
ntEP
for other approvals
otHEr
until MAStEr appears in the display.
and confirm with