Maintenance (cont'd)
Fuse Replacement
Electric shock hazard:
The fuses are an integral part of the overvoltage protection. When
fuse replacement is necessary, refer to "Specifications" for the
correct type, size and capacity. Using any other type of fuse will
void the overvoltage protection rating of the unit.
Failure to observe this warning could result in severe injury or
death.
1. Disconnect the unit from the circuit. Turn the unit OFF.
2. Remove the screw from the battery compartment cover. Remove
the battery compartment cover.
3. Remove the screws from the back cover. Remove the back cover.
4. Replace the fuse(s).
5. Replace the covers and screws.
Cleaning
Periodically wipe the case with a damp cloth and mild detergent; do
not use abrasives or solvents.
Greenlee Textron Inc. warrants to the original purchaser of these
goods for use that these products will be free from defects in
workmanship and material for their useful life, excepting normal
wear and abuse. This warranty is subject to the same terms and
conditions contained in Greenlee Textron Inc.'s standard one-year
limited warranty.