Pari BOY Pro Instrucciones De Uso página 139

Ocultar thumbs Ver también para BOY Pro:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 161
NEBULISER AND MASK
• Briefly rinse all parts used in running drinking water
beforehand.
• Place all disassembled com-
ponents in warm drinking water
with a little dishwashing liquid
for about 5 min.
• Rinse off all parts thoroughly in drinking water.
• Shake the water out of all of the parts.
Disinfecting
After cleaning, disinfect all of the disassembled parts (only
parts that have been cleaned can be disinfected effectively).
The recommended disinfection procedures are described be-
low. Descriptions of other validated disinfection procedures are
available from the manufacturer or dealer upon request.
The connection tube cannot be cleaned or disinfected. Re-
garding care of the connection tube, follow the instructions
in the corresponding section [see: Care of the connection
tube, page 34].
The elastic band cannot be disinfected.
CAUTION
Risk of infection due to moisture
Moisture encourages the growth of bacteria.
• Remove all parts from the pot or disinfector as soon as
the disinfection process is finished.
• Dry all parts.
– 31 –

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido