Guidance and manufacturer's declaration-electromagnetic immunity
for all eQuIPMent and systeMs
Guidance and manufacturer's declaration—electromagnetic emission
the Pulse Oximeter (18720) is intended for use in the electromagnetic environment specified below. the customer or the user
of the Pulse Oximeter (18720) should assure that it is used in such an environment
Immunity test
electrostatic
discharge (ESD) IEC
61000-4-2
Power frequency
(50Hz) magnetic field
ieC 61000-4-8
34
English
Iec60601
compliance level
test level
±6KV contact
±6KV contact
±8KV air
±8KV air
3a/m
3a/m
electromagnetic environment-guidance
floors should be wood, concrete or ceramic tile. if
floor are covered with synthetic material, the relative
humidity should be at least 30%.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commer-
cial or hospital environment