Ethernet Interface
Input
The scale will respond to various commands sent via the interface.
IP
Immediate Print of displayed weight (stable or unstable).
P
Print displayed weight (stable or unstable).
CP
Continuous Print.
SP
Print on Stability.
0S: Turn off "Stable Only" menu item and allow unstable print. 1S: Turn on "Stable
xS
Only" menu item and only print stable print.
xP
Interval Print x = Print Interval (1-3600 sec), 0P turns auto print OFF.
Z
Same as pressing Zero Key.
T
Same as pressing Tare Key.
Download Tare value in grams (positive values only).
xT
Sending 0T clears tare (if allowed).
PU
Print current unit: g, kg, lb, oz, lb:oz, t, c (custom).
xU
Set scale to unit x: 1=g, 2=kg, 3=lb, 4=oz, 5=lb:oz, 6=t.
xM
Set scale to mode x. M will scroll to next enabled mode.
PV
Version: print name, software revision and LFT ON (if LFT is set ON).
Enter Header line, where x = line number 1 to 5, "text" = header text up to 24
H x "text"
alphanumeric characters.
Enter Footer line, where x = line number 1 to 2, "text" = rear text up to 24 alphanumeric
F x "text"
characters.
Global reset to reset all menu settings to the original factory defaults
\EscR
(Escape key + 'R' Key).
Auto-Print Operation
Once Auto-Print is activated in the menu, the balance will send data as required. To temporarily stop
Auto-Printing, press the PRINT key. If there is data in the print buffer the printer will finish printing this
data. A second press will resume Auto-Printing.
(Note: In some modes, stopping or resuming Auto-Print must be done when the balance is not
reading "0".)
ACCESSORIES
For a complete listing of OHAUS printers and other accessories, contact OHAUS Corporation or visit
www.OHAUS.com