Diagnostics - Troubleshooting - Breakdowns - Tecno-gaz PURA Istrucciones Para El Uso

Tabla de contenido

Publicidad

11
DIAGNOSTICS – TROUBLESHOOTING - BREAKDOWNS
TROUBLESHOOTING TABLE
ATTENTION: WITH AUTOCLAVE SWITCH SWITCHED ON
ANOMALY
st
1
Autoclave is not
requesting water
nd
2
Supply tap closed
PURA™
produces no
water
rd
3
No power
st
1
Low waterworks
pressure
PURA™
produces
nd
very little
2
Defective or spent
water
osmotic membrane
The examples mentioned above are typical anomalies that may occur. Everything outside the third
column pertains to Tecno-Gaz S.p.a.
Contact the Service centre.
46
NORMAL SOLUTION (POSSIBLY
CAUSE
Check that the max level led of the
autoclave is off (see PAR 8).
Check that the tap supplying PURA™ is
not closed by mistake.
Check that you have inserted the
transformer plug correctly into the
autoclave socket and into the PURA™.
Check that the waterworks pressure is not
below 2 bar.
Nessuna azione
ADOPTED BY THE USER)
ENGLISH
SOLUTION BY
TECNO-GAZ
AUTHORIZED
TECHNICIANS
-
-
Check that the output
voltage of the
transformer is 24V DC.
Find the cause of the
low pressure and
eliminate it.
Replace the osmotic
membrane.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

247-s-2

Tabla de contenido