3.3 Assembly instructions: wall mount
100 - 240 V
Mounting parts
1. Push upper part of
examination lamp
onto the wall bracket.
2. Use caution when drilling!
Check the wall
for electric lines,
gas and water conduits.
3. Affix examination lamp
to the wall.
Adjust position of light
with both hands.
Electric supply
(EU / GB)
Turn ON - I -
Turn OFF - 0 -
16
HALOGEN or LED
4. CLEANING
Before cleaning, turn off the lamp. Unplug the power cord
and allow the lamp head to cool.
4.1 Floor stand: The surface of the floor stand can be kept clean by
wiping it gently with a damp cloth. Any common cleaning agent may
be used. For disinfection, we recommend a solution that is either dis-
solved in water or max. 20% alcohol.
4.2 Light body: The lamp is made with a high quality surface ma-
terial. The surface of the light can be easily kept clean by wiping it with
a damp cloth and any common cleaning agent. Only use disinfecting
agents with an alcohol content of less than 20%.
4.3 Examination lamp: 1. The examination lamp is resistant to
disinfectants, abrasion, urine and blood. 2. The examination lamp is
to be wiped with a damp cloth and alcohol-based surface disinfectant.
3. Clean dirty castors in order to avoid static charge. 4. Do not use
agents containing acids.
5. TECHNICAL DATA
Max. height: .................................................................. approx. 211 cm
Lamp arm pivoting range: .............................................. approx. 82 cm
Diameter of floor stand base: ......................................... approx. 70 cm
Weight: ............................................................................ approx. 4.4 kg
Supply voltage: .............................................................. 100 - 240 Volt
Halogen: ............................................................ 12 V / 35 W / 5 A max.
Life of the halogen bulb: ....................................................... ≥ 1.000 h
LED: ................................................................ 12V / 6,5W / 0,8 A max.
Life of the LED lamp: .................................................... approx. 50.000h
Beam angle HL / LED: ................................................................ 8° / 24°
3.4 Replacing the bulb
1. Disconnect the
power plug.
2. Set switch to - 0 -.
3. Unscrew the ring.
4. Remove defective
bulb only after it has
cooled down.
5. Insert new bulb.
6. Screw the ring back on.
Colour temperature ca.: .....................................HL: 3200K/LED: 4000K
Illumination intensity (HL) at a distance of 30 cm .................. 44.000 Lux
Illumination intensity (LED) at a distance of 30 cm ................. 31.000 Lux
Operating temperature range: ............................... + 10° C to + 40° C
Relative humidity ........................................................... 30 % to 75 %
Air pressure: ........................................................... 700 hPa to 1060 hPa
Storage temperature range: ................................... - 10° C bis + 50° C
Light spot size (at a working distance of 0.5m) ...................... 100 mm
Note: The technical data are subject to certain fluctuations, especially
with regard to the product life, color temperature and illumination
intensity. The floor stand model should be placed on level surfaces.
Information: The examination lamp is not water proof! Na-
tional board requirements (standards and directives) for hygiene and
disinfection must be followed.
6. MAINTENANCE
After being used for a longer period of time, the KaWe MASTERLIGHT
HL / LED lamp screws must be tightened. The lamp is to be serviced and
checked at least every two years.
Information: The halogen lamp can become hot during use!
When servicing and or checking the unit, turn it off and disconnect the
plug. Insure that the lamp cannot be turned back on.
6.1 Regular maintenance: The structural system of the lamp is
to be checked/serviced every two years: 1. Cracks in plastic parts
2. Functionality test 3. Electrical safety test 4. Structural system check.
17