Auto hold function
For weights above 2 kg, the HOLD function
is activated automatically. The display flash-
es until a stable weight has been measured.
The display is then frozen until the next
weighing operation takes place. HOLD ap-
pears in the display.
Automatic shut-down/power-saving
mode
To save power, the scale switches itself off
after a short time if there is no load on it.
Switching the weight display
Switch on the scale. Weight is initially dis-
played in the last setting used, either in kilo-
grammes or pounds.
To change the weight display, press the kg/
lbs key until the display changes.
Taring weight display (TARE)
The tare function is useful if an additional
weight (for example a cloth or pad on the
weighing surface) is to be ignored in the
weighing result. Proceed as follows:
press the TARE key while the additional
weight is on the scale and wait until
appears in the display.
Important!
The maximum additional weight is 10 kg.
The maximum weight which can be dis-
played is reduced by the weight of the ob-
jects already placed on the scale.
You can now perform as many weighing
operations as you like with this additional
weight.
You can switch off TARE by pressing the
TARE key again or by switching off the
scale.
Cleaning
Clean the weighing tray and the housing as
required using a domestic cleaning agent or
a commercially-available disinfectant. Fol-
low the manufacturer's instructions.
What do I do if ...
... no weight is displayed when there is
a weight on the scale?
Is the scale switched on?
Check the batteries.
...
0. 0 00
does not appear before weigh-
ing?
0. 0 00
Switch the scale off and back on again,
making sure there is no load on the scale.
... one segment comes on permanently
or not at all?
This segment is faulty. Inform Servicing.
...
appears in the display?
Battery power is dropping. Change the bat-
teries in the next few days.
batt
...
appears in the display?
The batteries are completely discharged.
Insert new batteries.
St0p
...
appears in the display?
The maximum load has been exceeded.
...
appears in the display?
The ambient temperature of the scale is too
high or too low. Place the scale in an ambi-
ent temperature between +10 ˚C and
+40 ˚C. Wait about 15 minutes until the
scale has acclimatised to the ambient tem-
perature and then weigh again.
Er:011:
...
appears in the display?
The scale is overloaded or the load is une-
venly distributed. Take the load off the scale
or distribute weight more evenly and restart
scale.
...
Er:012:
appears in the display?
The scale was switched on with too great a
load on it. Remove load and restart.
7