Operating Instructions; Calibration - SKF TMEH 1 Instrucciones De Uso

Ocultar thumbs Ver también para TMEH 1:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 29
3.

OPERATING INSTRUCTIONS

3.1

Calibration

Before calibration it is important to
ensure the sensing cell is clean
and dry, as moisture and
contamination will adversely
effect the results. Testing should
generally be undertaken with the
unit switched in the HI position.
The Green/Red scale is mainly for
use when testing engine oils.
Step 1
Half fill the sensing cell with clean
oil (ensure that the whole surface
of the sensor is well covered with
oil). Oil of the same brand as in
the system being evaluated,
should be used for calibration.
Step 2
Run the rotary switch from the
OFF position to the HI position,
the display will complete a full
sweep and return to the zero
position. The display will show
with the CAL symbol flashing.
If during testing the CAL graphic
flashes the unit will require re-
calibration.
SKF TMEH 1
OilCheck
7

Publicidad

Tabla de contenido
loading

Tabla de contenido