KPS Multicheck6010 Manual De Usuario página 65

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Multicheck6010
Step 4
Ensure that the item for testing is disconnected from any voltage source and it has
been fully discharged. Connect the test cables to the item under test.
Follow the connection diagrams shown in figure 4.6 to perform a R Low resistance
measurement.
Step 5
Check for any warnings and the online voltage/terminal monitor on the display before
starting the measurement. If everything is ok and the
After performing the measurement, the results appear on the display together with the
After performing the measurement, the results appear on the display together with the
or
indication (if applicable).
Figure 4 7: Examples of R Low resistance measurement results
Displayed results:
R
Main LowΩ resistance result (average of R+ and R- results)
R+
LowΩ resistance sub-result with positive voltage at L terminal
R-
LowΩ resistance sub-result with positive voltage at N terminal
Warnings:
• Low-value resistance measurements should only be performed on de-energized objects!
• Parallel impedances or transient currents may influence test results.
Note:
• If voltage between test terminals is higher than 10 V the R Low measurement will not be
performed.
Figure 4 6: Connection of universal test cable
- 65 -
is shown, press the TEST key.
EN