EMC Technical Information
Table 1
The "Ease-It™" is intended for use in the electromagnetic environment specified below. The customer or the user of the "Ease-It™"
should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR11
IEC 60601-1-2
EN55011
RF emissions
CISPR11
IEC 60601-1-2
EN55011
Harmonic emissions
IEC 60601-1-2
IEC 61000-3-2
Voltage fluctuations/
flicker emissions IEC
60601-1-2
IEC 61000-3-3
Table 2
The "Ease-It™" is intended for use in the electromagnetic environment specified below. The customer or the user of the "Ease-It™"
should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 60601-1-2
IEC61000-4-2
Electrical fast transient/
burst
IEC 60601-1-
2 IEC61000-
4-4
Surge
IEC 60601-1-
2 IEC61000-
4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines.
IEC 60601-1-2
IEC61000-4-
11
Power frequency (50-60Hz)
magnetic field.
IEC 60601-1-
2 IEC61000-
4-8
Electromagnetic emissions
Compliance
Group 1
Class B
Class A
Complies
Electromagnetic immunity
IEC 60601-1
Test level
±6KV contact
±8KV air
±15KV air
±2KV for power supply
lines
±1KV for input/output lines
±1KV differential mode
<5% UT for 0.5 cycle
40%UT for 5 cycles
70%UT for 25 cycles
<5%UT for 250 cycles
30A/m
Electromagnetic environment guidance
The device uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The device is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Compliance level
±6KV contact
±8KV air
±15KV air
±2KV for power supply
lines
±1KV for input/output lines
±1KV differential mode
<5% UT for 0.5 cycle
40%UT for 5 cycles
70%UT for 25 cycles
<5%UT for 250 cycles
30A/m
Electromagnetic environment
guidance
Floors should be wood, concrete or
ceramic tile.
If floors are covered with synthetic
material, the relative humidity should
be at least 40%.
Mains power supply quality should
be that of typical residential area.
Mains power supply quality should
be that of typical residential area.
Power frequency magnetic fields
should be that of typical residential
area.